Appendix A: Emc Information - GCE Healthcare M50 Manual De Usuario

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 48
If you have any questions regarding the guidance and declarations listed below, please contact your home
care provider.
Guidance and Manufacturer's Declaration - Electromagnetic Emissions: This device is intended for use in
the electromagnetic environment specifi ed below. The user of this device should make sure that it is used
in such an environment.
EMISSIONS TEST
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fl uctuations/Flicker
emissions IEC 61000-3-3
Guidance and Manufacturer's Declaration - Electromagnetic Immunity: This device is intended for use in
the electromagnetic environment specifi ed below. The user of this device should make sure that it is used
in such an environment.
IMMUNITY TEST
Electrostatic
Discharge (ESD)
IEC 61000-4-2
Electrical Fast
Transient/Burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
Power frequency
(50/60 Hz) magnetic
fi eld IEC 61000-4-8

APPENDIX A: EMC INFORMATION

COMPLIANCE
Group 1
Class B
Class A
Complies
IEC 60601 TEST LEVEL
±15kV air
±8kV contact
±2kV for power
supply lines
±1kV for input-output
lines
±1 kV diff erential
mode
±2 kV common
mode
<5% UT (>95% dip in
UT) for 0.5 cycle 40%
UT (60% dip in UT) for 5
cycles 70% UT
(30% dip in UT) for 25
cycles <5% UT
(>95% dip in UT) for 5
seconds
30 A/m
11/135
ELECTROMAGNETIC
ENVIRONMENT-GUIDANCE
This device uses RF energy only
for its internal function. Therefo-
re, its RF emissions are very low
and are not likely to cause any
interference in nearby electronic
equipment.
The device is suitable for use
in all establishments, including
domestic establishments and
those directly connected to the
public low-voltage power supply
network.
COMPLIANCE LEVEL
±15kV air
±8kV contact
±2 kV for supply
mains
±1 kV for input/output
lines
±1 kV diff erential
mode
±2 kV common
mode
<5% UT (>95% dip in
UT) for 0.5 cycle 40%
UT (60% dip in UT)
for5 cycles 70% UT
(30% dip in UT) for 25
cycles <5% UT
(>95% dip in UT) for 5
seconds
30 A/m
ELECTROMAGNETIC
ENVIRONMENT-
GUIDANCE
Floors should be wood,
concrete or ceramic tile.
If fl oors are covered with
synthetic material, the rela-
tive humidity should be at
least 30%.
Mains power quality
should be that of a ty-
pical home or hospital
environment.
Mains power quality
should be that of a ty-
pical home or hospital
environment.
Mains power quality
should be that of a ty-
pical home or hospital
environment.
If the user of the de-
vice requires continued
operation during power
mains interruptions, it is
recommended that the
device be powered from
an uninterruptible power
supply or battery.
Power frequency magnetic
fi elds should be at levels
characteristic of a typical
location in a typical hospi-
tal or home environment.
EN

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

M50

Tabla de contenido