Warning
Turning the torque wrench further clockwise will
damage the handpiece.
You must use the torque base when assembling
and dissembling the tip from the handpiece.
Failure to use a torque base could result in
a twisted transducer within the handpiece
assembly, potentially damaging the transducer,
and leading to handpiece failure.
Symptoms of an over-torqued handpiece include:
Difficulty inserting the aspiration tubing
into the nosecone
Inadequate aspiration
Lower than expected handpiece power
output
6. Carefully remove the wrench from the tip.
7. Remove the handpiece from the tip torquing
base.
8. Locate the o-rings for the handpiece you are
assembling:
Large o-ring
Small o-ring
9. Slide the larger o-ring over the connecting body
and into the groove in the neck of the handpiece.
Slide the smaller o-ring into the groove in the
metal connecting body.
10. Put the assembled handpiece, nosecone, and one
flue with one tip cleaner into the sterilizer case.
Warning
The handpiece and handpiece accessories must be
sterile before use.
23 kHz
36 kHz
black
white
green
blue
Completing Handpiece Assembly
Refer to the CUSA Excel System User's Guide for
instructions on how to attach the nosecone, flue
and manifold tubing to the handpiece.
Testing the Handpiece
Once you have completed system setup, test the
handpiece to ensure that you have assembled
the tip correctly. Refer to the CUSA Excel System
User's Guide for instructions on how to test the
handpiece.
Disassembly
When cleaning handpieces, be sure to wear
protective clothing, gloves, and safety glasses
according to your hospital's policy for the cleaning
process.
1. Using a soft cloth moistened with a germicidal
solution, wipe the handpiece to remove the gross
contaminants.
2. Remove the flue, if not already removed.
3. Remove the nosecone:
a. Release the nosecone by twisting it
counterclockwise.
b. Pull the nosecone away from the handpiece.
DO NOT DISCARD A STANDARD NOSECONE.
EN