Storz ENDOMAT UP 210 Manual De Instrucciones página 93

Ocultar thumbs Ver también para ENDOMAT UP 210:
Tabla de contenido

Publicidad

85
Hinweise zur
elektromagnetischen
Verträglichkeit (EMV)
Recommended separation distances between portable and mobile RF communications equipment and the ENDOMAT
The ENDOMAT
SELECT UP 210 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer
®
or user of the ENDOMAT
SELECT UP 210 can help prevent electromagnetic interference by maintaining a minimum distance between portable and
®
mobile RF communications equipment (transmitters) and the ENDOMAT
power of the communications equipment.
Rated maximum output power
of transmitter
[W]
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance D in meters (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
Note: At 80 MHz and 800 MHz, the separation distance for higher frequency range applies.
Note: These guidelines may not apply in all situations. The propagation of electromagnetic quantities is affected by absorptions and reflections of structures,
objects and people.
HF surgical instrument compatibility
This equipment has been evaluated for compatibility
with high-frequency surgical equipment. It has been
tested according to IEC 60601-2-2 Annex BB.
Electromagnetic
Compatibility
(EMC) Information
Table 5
SELECT UP 210 as recommended below, according to the maximum output
®
Separation distance d [m] as a function of the transmitter frequency
150 kHz to 80 MHz
80 MHz to 800 MHz
f
0.12
0.38
1.2
3.8
12
Indicaciones sobre
compatibilidad
electromagnética (CEM)
800 MHz to 2.5 GHz
f
h
0.12
0.38
1.2
3.8
12
SELECT UP 210
®
0.23
0.73
2.3
7.3
23

Publicidad

Tabla de contenido
loading