Safety Information
If you use other medical devices or wear implantable medical devices such as
defibrillators or pacemakers and are concerned that your Miracle-EarLINK
Remote Mic AV might cause interference with your medical device, please
contact your physician or the manufacturer of your medical device for
information about the risk of disturbance. The Miracle-EarLINK
AV should not be used during an MRI procedure or in a hyperbaric chamber, or
other oxygen-rich environments. The Miracle-EarLINK
formally certified to operate in explosive atmospheres such as may be found in
coal mines or certain chemical factories.
Use on Aircrafts
The Miracle-EarLINK
Remote Mic AV should not be used on aircrafts unless
™
specifically permitted by the flight personnel.
International Use
The Miracle-EarLINK
Remote Mic AV is approved to operate at radio
™
frequencies that are specific to your country or region and might not be
approved for use outside your country or region.
Be aware that operation during international travel may cause interference with
other electronic devices, or other electronic devices may cause interference
with your Miracle-EarLINK
™
Remote Mic AV.
Your Miracle-EarLINK
95% rH
45°C
designed to operate and be stored and
10% rH
transported within the temperature range of
-10°C
-10˚C (14˚F) to 45˚C (113˚F) and relative humidity
range of 10%-95%.
We are required by regulations to provide the following warnings:
WARNING:
Use of the Miracle-EarLINK
to other electronic equipment should be avoided because it could result
in improper performance. If such use is necessary, note as to whether your
hearing aids and the other equipment are operating normally.
WARNING:
Use of accessories, components or replacement parts other
than those provided by the manufacturer of the Miracle-EarLINK
Mic AV could result in increased electromagnetic emissions and decreased
electromagnetic immunity and could result in degradation of performance.
WARNING:
If Portable Radio Frequency communications equipment is
used closer than 30 cm (12 inches) from your Miracle-EarLINK
degradation of the performance of your hearing aid could result. If this occurs,
move away from the communications equipment.
20
™
Remote Mic
™
™
Remote Mic AV is not
™
Remote Mic AV is
Remote Mic AV directly next
™
Remote
™
™
Remote Mic AV,
Safety Information
Although the Miracle-EarLINK
™
Remote Mic AV is not formally certified (IP
rated) for protection against dust or water ingress, per IEC 60529, it has been
tested to the IP 5X level. This means that the Miracle-EarLINK
is protected from dust.
The battery in your Miracle-EarLINK
™
attempt to replace the battery. The Miracle-EarLINK
contain any user-serviceable parts. Do not open or modify it.
Do not connect the Miracle-EarLINK
™
described in this manual. The expected service life of the Miracle-EarLINK
Remote Mic AV is 3 years.
Wireless Technical Description
Miracle-EarLINK
Remote Mic AV operates in the 2.4- 2.4835 GHz band with
™
a maximum effective radiated power of 8 dBm with transmission modulation
type of GFSK with 2 MHz bandwidth. The receiver section of the radio has a
bandwidth of 2 MHz.
The Miracle-EarLINK
Remote Mic AV has been tested to, and has passed, the
™
following emissions and immunity tests:
• IEC 60601-1-2 radiated and conducted emissions requirements for a
Group 1 Class B device as stated in CISPR 11.
• Harmonic distortion and voltage fluctuations affecting the power input
source as stated in Table 2 of IEC 60601-1-2.
• RF radiated immunity at a field level of 10 V/m between 80 MHz and 2.7
GHz as well as higher field levels from communications devices as stated
in Table 9 of IEC 60601-1-2.
• Immunity to power frequency magnetic fields at a field level of 30 A/m
as well as proximity magnetic fields as defined in table 11 of 60601-1-2.
• Immunity to ESD levels of +/- 8 kV conducted discharge and +/- 15 kV
air discharge.
• Immunity to electrical fast transients on the power input at a level of +/-
2 kV at a 100 Hz repetition rate.
• Immunity to surges on the power input of +/- 1 kV line to line.
• Immunity to conducted disturbances induced by RF fields on the power
input as stated in Table 6 of IEC 60601-1-2.
• Immunity to voltage dips and interruptions on the power input as stated
in Table 6 of IEC 60601-1-2.
Remote Mic AV
™
Remote Mic AV is not replaceable. Do not
Remote Mic AV does not
™
Remote Mic AV to any devices not
™
21