Pump Maintenance - Hach AS950 AWRS Manual De Instrucciones

Ocultar thumbs Ver también para AS950 AWRS:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

Figure 13 Desiccant cartridge
1 Plug
2 Desiccant tube

7.3 Pump maintenance

Pinch hazard. Remove power from the instrument before maintenance or service activities are done.
7.3.1 Replace the pump tubing
Use of tubing other than that supplied by the manufacturer can cause excessive wear on mechanical parts and/or
poor pump performance.
Examine the pump tubing for wear where the rollers rub against the tubing. Replace the tubing when
the tubing shows signs of wear.
Pre-requisites:
• Pump tubing—pre-cut or bulk 4.6 m or 15.2 m (15 ft or 50 ft)
1. Disconnect the power to the controller.
2. If the bulk tubing is used, cut the tubing and add alignment dots. Refer to
3. Remove the pump tubing as shown in the illustrated steps that follow.
4. Clean the silicone residue from the interior of the pump housing and from the rollers.
5. Install the new pump tubing as shown in the illustrated steps that follow.
3 O-ring
4 Desiccant window
C A U T I O N
N O T I C E
Figure
14.
English 21

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

R600a

Tabla de contenido