Hach AS950 AWRS Manual De Instrucciones página 4

Ocultar thumbs Ver también para AS950 AWRS:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

Figure 1 AWRS
1 Cover latch
2 AWRS
3 Access cover
4 Liquid detector
Section 3 Specifications
Specifications are subject to change without notice.
3.1 All Weather Refrigerated Sampler (AWRS)
Specification
Dimensions (W x D x H)
Weight
Power requirements (includes
compressor)
Overload protection
Compressor
Operating temperature
Storage temperature
Relative humidity
Overvoltage category
Pollution degree
Protection class
Altitude
1
Refer to
Figure 2
4 English
5 Non-contacting liquid detector
6 Controller cover
7 Compartment heater option
8 Pump
AWRS
1
76 x 81 x 130 cm (30 x 32 x 51 in.)
86 kg (190 lb)
230 VAC, 50 Hz, 2.7 A or 4.1 A with controller compartment heater
230 VAC: 5.0 A circuit breaker
230 VAC, 50 Hz, 3.0 A peak start current
0 to 50 °C (32 to 122 °F); with AC battery backup: 0 to 40 ºC (32 to 104 ºF);
with controller compartment heater: –40 to 50 °C (–40 to 122 °F); with
controller compartment heater and AC battery backup: –15 to 40 ºC (5 to
104 ºF)
–30 to 60 °C (–22 to 140 °F)
0 to 95%
II
2
I
2000 m (6562 ft) maximum
on page 7 for the sampler dimensions.
9 Door latch
10 Controller
11 Refrigerated cabinet
12 Bottle tray

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

R600a

Tabla de contenido