Dusthunter Sb; Rated Data For Ambient Conditions; Rated Data For Devices With Ingress Protection; Supplements To Technical Data Section - SICK DUSTHUNTER C200 Instrucciones De Operación

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

8

DUSTHUNTER SB

8
DUSTHUNTER SB
8.1

Rated data for ambient conditions

Use inside or outside buildings
Altitude
Ambient temperature
Max. relative humidity
Line voltage fluctuations
Overvoltage category
Wet environment
Degree of contamination
[1] Recommendation: Use optional weather protection hoods for device components.
EN
8.2

Rated data for devices with ingress protection

[1]
Enclosure rating
[1] Fitting position of MCU control unit: Cables must be routed at the bottom to ensure IP protection
8.3

Supplements to Technical Data Section

8.3.1
Electrical data
Battery type:
Chemical system:
Device component:
22
A D D E N D U M T O O P E R A T I N G I N S T R U C T I O N S | Dust measuring devices and Combined probe
SB30
[1]
Indoor and outdoor
Max. 2000 m (above sea level)
‒20 ... +55 °C
Option MCU-P: ‒40 ... +45 °C for purge air intake temperature
<95% non-condensing
Sender/receiver unit: 24 V DC, ±10%
Option MCU: 90 ... 250 V AC, 47 ... 63 Hz; opt. 24 V DC, ±2 V
Option purge air unit SLV4 (2BH13):
3-phase: 200 ... 240 V / 345 ... 415 V, 50 Hz
3-phase: 220 ... 275 V/ 380 ... 480 V, 60 Hz
Overvoltage category II
Not relevant
Degree of contamination 2
SB30
IP66
Option purge air unit SLV4: IP54
Power connection of MCU control unit
Conductor
Power cable
Battery
SB30 without control unit
Button cell BR2032
Lithium poly-carbon monofluoride
Sender/receiver unit
SB50/100
[1]
Indoor and outdoor
Max. 2000 m (above sea level)
‒20 ... +55 °C
Option MCU-P: ‒40 ... +45 °C for purge air intake temperature
<95% non-condensing
Sender/receiver unit: 90 ... 250 V AC, 47 ... 63 Hz;
opt. 24 V DC, ±2 V
Option purge air unit SLV4 (2BH13):
3-phase: 200 ... 240 V / 345 ... 415 V, 50 Hz
3-phase: 220 ... 275 V / 380 ... 480 V, 60 Hz
Option DUSTHUNTER SB-FSS: 19 ... 28 V DC
Overvoltage category II
Not relevant
Degree of contamination 2
SB50/100
IP66
Option purge air unit: IP54
Cross-section
2
2
1.5 mm
... 4 mm
SB30/50/100
Button cell BR2032
Lithium poly-carbon monofluoride
MCU control unit
8017305/AE00/V1-0/2022-09| SICK
Subject to change without notice