DUSTHUNTER SP, Combiprobe CP100

9.1

Rated data for ambient conditions

SP30
Use inside or outside buildings
Indoor and outdoor
Altitude
Max. 2000 m (above sea level)
Ambient temperature
‒20 ... +55 °C
Option MCU-P: ‒40 ... +45 °C for purge air
intake temperature
Max. relative humidity
<95% non-condensing
Line voltage fluctuations
Sender/receiver unit: 24 V DC, ±10%
Option MCU: 90 ... 250 V AC, 47 ... 63 Hz; opt.
24 V DC, ±2 V
Option purge air unit SLV4 (2BH13):
3-phase: 200 ... 240 V / 345 ... 415 V, 50 Hz
3-phase: 220 ... 275 V / 380 ... 480 V, 60 Hz
Overvoltage category
Overvoltage category II
Wet environment
Not relevant
Degree of contamination
Degree of contamination 2
[1] Recommendation: Use optional weather protection hoods for device components.
9.2

Rated data for devices with ingress protection

SP30
Enclosure rating
[1]
Sender/receiver unit DUSTHUNTER SP30
without integrated purge air unit: IP66
Sender/receiver unit DUSTHUNTER SP30
with integrated purge air unit: IP54
[1] Fitting position of MCU/MCUDH Ex-3K control unit: Cables must be routed at the bottom to ensure IP protection
8017305/AE00/V1-0/2022-09| SICK
Subject to change without notice
SP100, SP100 Ex-2K, SP100 Ex-3K
[1]
Indoor and outdoor
Max. 2000 m (above sea level)
DUSTHUNTER SP100:
Sender/receiver unit, MCU-N: ‒20 ... +55 °C
Option MCU-P: ‒40 ... +45 °C for purge air
intake temperature
DUSTHUNTER SP100 Ex-3K
DUSTHUNTER SP-TxxxxEx-
3K: −15 °C ... +60 °C
MCU: −40 °C ... +60 °C
MCUDH Ex-3K NSxxx: −25 °C ... +50 °C
MCUDH Ex-3K N2xxx: −40 °C ... +60 °C
DUSTHUNTER SP100 Ex-2K
DUSTHUNTER SP-TxxxxEx-2K:
−40 °C ... +60 °C
MCUDH Ex-3K NSxxx: −25 °C ... +50 °C
MCUDH Ex-3K N2xxx: −40 °C ... +60 °C
<95% non-condensing
DUSTHUNTER SP100
MCU: 90 ... 250 V AC, 47 ... 63 Hz;
opt. 24 V DC, ±2 V
Option purge air unit SLV4 (2BH13):
3-phase: 200 ... 240 V/345 ... 415 V, 50 Hz
3-phase: 220 ... 275 V/380 ... 480 V, 60 Hz
DUSTHUNTER SP100 Ex-3K
DUSTHUNTER SP-TxxxxEx-3K: 24 V DC
MCU: 90 ... 250 V AC, 47 ... 63 Hz;
opt. 24 V DC
MCUDH Ex-3K: 90 ... 250 V AC, 47 ... 63 Hz;
opt. 24 V DC, ±2 V
DUSTHUNTER SP100 Ex-2K
DUSTHUNTER SP-TxxxxEx-2K: 24 V
MCUDH Ex-3K: 90 ... 250 V AC, 47 ... 63 Hz;
opt. 24 V DC, ±2 V
Overvoltage category II
Not relevant
Degree of contamination 2
SP100, SP100 Ex-2K, SP100 Ex-3K
DUSTHUNTER SP100: IP66
Option purge air unit: IP54
DUSTHUNTER SP100 Ex-3K/-2K: IP66
MCU: IP65
MCUDH Ex-3K: IP65
A D D E N D U M T O O P E R A T I N G I N S T R U C T I O N S | Dust measuring devices and Combined probe
DUSTHUNTER SP, COMBIPROBE CP100
CP100
[1]
Indoor and outdoor
Max. 2000 m (above sea level)
‒20 ... +55 °C
Option MCU-P: ‒40 ... +45 °C for
purge air intake temperature
<95% non-condensing
DUSTHUNTER SP100/
FLOWSIC100 PR:
90 ... 250 V AC, 47 ... 63 Hz;
opt. 24 V DC, ±2 V
Pressure, temperature sensor:
10 ... 30 V DC
Overvoltage category II
Not relevant
Degree of contamination 2
CP100
DUSTHUNTER SP100: IP66
FLOWSIC100 PR: IP65
Pressure sensor: IP54
Temperature sensor: IP54