8.9
Installing fluorescence filter (B-510LD series)
1. Disconnect the power supply plug from the fluorescence il-
luminator.
2. Open the side cover of the illuminator, by unscrewing the
side screws ①. (Fig. 43)
•
It could be helpful to remove the observation head.
•
The cubes are mounted on the opposite side of the cover:
opening the left cover acts on the right side of the slider and
vice versa.
3. Open the top door of the fluorescence illuminator by un-
screwing the four screws ② and release the cover. (Fig. 44)
4. Loosen the front locking screw ③ on the fluorescence cube
slider, corresponding to the fluorescence cube to be installed
or replaced. (Fig. 45)
5. Insert the fluorescence cube into the dovetail ④ of the cube
slider and move it to the click position. (Fig. 46)
6. Tighten the locking screw ③. (Fig. 45)
Page 35