5. Open the lamp housing using the door lock screw ⑤ and
remove the lamp holder. (Fig. 19)
6. Remove the plastic block ⑥ from the lamp holder (or the
exhausted lamp in case of replacement) by loosening the two
locking screws ⑦. (Fig. 20)
7. Insert the mercury bulb ⑧ (respect the polarity of the bulb),
tighten the locking screws and refit the lamp holder into the
lamp housing. (Fig. 21)
Disconnect
installing or replacing the bulb.
The bulb has an anode and a cathode of
different sizes. Respect the polarity during
assembly, respecting the bulb dimensions.
Do not touch the bulb of the lamp with bare
hands to leave no traces of grease on the
bulb. If this happens, clean the bulb with a
soft cloth before turning on the lamp.
The bulb has an average life of about 200-250
hours: a time counter and a voltage indicator
are shown on the bulb power supply. Replace
the bulb when the hour count exceeds 250 or
if the voltage drops below 4.5A.
During use, the bulb, the lamp housing and the surrounding environment become hot.
Before replacing the lamp, switch off the power supply, disconnect all cables and wait for the bulb
and the lamp housing to cool.
After switching on the bulb, wait at least 10-15 minutes before switching it off.
After switching off the bulb, wait for 5-10 minutes before switching it on again so that the mercury
vapors have time to condense.
The bulb contains ultraviolet radiation that could be harmful to eyes and skin. Always look at the
bulb arc through the provided orange screen.