7. Connect the plug of the fluorescence cube ⑤ in one of the
free connectors ⑥ to power the LED. (Fig. 47)
8. Apply the adhesive marker ⑦ for the fluorescence cube on
the illuminator. (Fig. 48)
9. Close the top door.
10. Close the side cover.
11. Connect the power supply.
12. Start working.

8.10 Replacing fluorescence filter (B-510LD series)

1. Disconnect the power supply plug from the fluorescence il-
luminator.
2. Open the side cover of the illuminator, by unscrewing the
side screws ①. (Fig. 43)
It could be helpful to remove the observation head.
The cubes are mounted on the opposite side of the cover:
opening the left cover acts on the right side of the slide and
vice versa.
3. Open the top door of the fluorescence illuminator by un-
screwing the four screws ② and release the cover. (Fig. 44)
4. Loosen the front locking screw ③ on the fluorescence cube
slider. (Fig. 45)
5. Disconnect the plug ⑤ related to the cube you want to re-
place. (Fig. 47)
6. Remove the fluorescence cube from the dovetail ④ of the
cube slider.
7. Repeat steps 4. to 9. of paragraph 8.9 to install a new fluo-
rescence cube.