Use Of The Microscope (B-510Fl); Use Of The Shutter (B-510Fl Only) - Optika B-510 Serie Manual De Instrucciones

Ocultar thumbs Ver también para B-510 Serie:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 122

15.2 Use of the microscope (B-510FL)

1. Turn on the power supply ① for the mercury bulb and wait 5
minutes for the arc to stabilize. (Fig. 93)
2. Move the filter selector ② to one of the four available positions
until the click stop. (Fig. 94)
The microscope has a 4-position filter holder slider. The
positions 1 and 2 are empty to house additional filters,
position 3 houses a B filter and position 4 a G filter.
F I L T E R
EXCITATION
NAME
FILTER
B*
460-490 nm
G*
510-550 nm
UV**
325-375 nm
V**
390-420 nm
(*)
Already installed in the microscope
(**) Optional

15.2.1 Use of the shutter (B-510FL only)

The microscope is equipped with a shutter ③ located on
the right side of the fluorescent illuminator. (Fig. 95)
1. Close the shutter in order to interrupt the observation for a
limited time and not subjecting the sample to unnecessary
lighting in the period in which it is not observed. (Switching
off and switching on frequently the HBO lamp considerably
reduces its duration).
D I C H R O I C
BARRIER
MIRROR
FILTER
505 nm
515LP nm
570 nm
575LP nm
415 nm
435LP nm
440 nm
455LP nm
FITC
GFP
Rhodamine
TRITC
Propidium Iodide
DAPI
Hoechst
Coumarin
Pacific Blue
Page 51
APPLICATIONS
F
F ig. 93
ig. 93
F ig. 94
F
ig. 94
F ig. 95
F
ig. 95

Publicidad

Tabla de contenido
loading

Tabla de contenido