IV. WHEELCHAIR USE & CONFIGURATION
BB. Power Tilt (Continued)
WARNING: In the unlikely event of battery acid leakage and you get acid in your eyes or
skin - rinse thoroughly with water and contact your local healthcare professional for
further consultation.
•
CHOKING HAZARD - The system is equipped with a charger connector plug to seal
the connector hole, be aware that this plug can be removed and swallowed by a
small child.
CAUTION: Do not remove interior or posterior tilt stops (Fig. 44:A). Removing tilt stops
may cause the system to to operate beyond its design limits causing system or
component failure.
Transfers
WARNING: Avoid the switch system and other adaptive switches when transferring to
prevent unintentional chair tilting. Failure to comply with the instruction above could result
in serious injury or death.
Cleaning and Disinfection
The RCS system is designed to fulfill IPx4 (or IPx6 as an option), which allows cleaning with
water. For disinfection, it is recommended to use soap or equal and clean with a paper
towel/rag.
Controls: 1-4 channel individual actuators
Current Limit Control: Default 8 A
System Power: 24VDC
•2 Ah (sealed lead accumulators)
•Ventilated housing
Charger: Power supply 120 VAC - 230 VAC
Protection Class: IPX4
Color: Black and orange
Fig. 44
Batteries:
A
Standard Features
Weight: 1.9 kg (4.19 lbs) (including batteries)
Operating Temperature: -15°C to +50°C (5°F to
Storage Temperature: -40°C to +60°C (-40°F to
Flammability Rating: UL94V-0
47
A
EMC Conformity:
•EN 61000-6-2 •EN 60601-1-2
•EN 61000-6-3 •ISO 7176-21
122°F)
140°F)
Soft Start / Stop
Single Fault Protection