28
Anhang
Hinweise zur elektromagnetischen
Verträglichkeit (EMV) für
Ladegerät Modell 094129/11301 DG
Guidance and Manufacturer's Declaration – Electromagnetic Interference Immunity
The charger model 094129/11301 DG is designed for operation in an electromagnetic environment as defined below.
The user of the charger model 094129/11301 DG should make sure that it is used in such an environment.
Interference immunity test
EN°IEC°60601 test level
Electrostatic Discharge (ESD)
IEC 61000-4-2
± 6 kV Contact discharge
± 8 kV Air discharge
Rapid transient electric
interferences /bursts
± 2 kV for mains lines
according to
± 1 kV for input and
IEC 61000-4-4
output lines
Surges
according to
± 1 kV Push-pull voltage
IEC 61000-4-5
± 2 kV Common-mode
voltage
Voltage dips, short-time
interruptions and fluctuations
<5% U
of the supply voltage
(>95% dip in U
according to IEC 61000-4-11
for ½ cycle
40% U
(60% dip in U
for 5 cycles
70% U
(30% dip in U
for 25 cycles
<5% U
(>95% dip in U
for 5 seconds
Magnetic field at the supply
frequency
3 A/m
(50/60 Hz)
according to IEC 61000-4-8
* Note: U
is the mains alternating voltage before applying the test levels.
T
Appendix
Information on the electromagnetic
compatibility (EMC) of
charger model 094129/11301 DG
Table 202
Conformity level
Conforms
± 6 kV Contact discharge
± 8 kV Air discharge
Conforms
± 2 kV for mains lines
± 1 kV for input and
output liness
Conforms
± 1 kV Push-pull voltage
± 2 kV Common-mode
voltage
Conforms
*
<5% U
*
T
T
)
(>95% dip in U
T
for ½ cycle
Conforms
40% U
T
T
)
(60% dip in U
T
T
for 5 cycles
Conforms
70% U
T
T
)
(30% dip in U
T
T
for 25 cycles
Conforms
<5% U
T
T
)
(>95% dip in U
T
for 5 seconds
Conforms
3 A/m
Anexo
Indicaciones relativas a la compatibilidad
electromagnética (CEM) del cargador,
modelo 094129/11301 DG
Electromagnetic environment – guidelines
Floors should be made of wood or concrete or
provided with ceramic tiles. If the floor is provided with
synthetic material, the relative humidity must be at
least 30%.
The quality of the supply voltage
should be that of a typical business
or hospital environment.
The quality of the supply voltage
should be that of a typical business
or hospital environment.
The quality of the supply voltage
should be that of a typical business
)
or hospital environment.
T
If the user of the device requests continued functioning
even when interruptions of the power supply occur, it
)
is recommended to supply the device from a power
supply that is free of interruptions.
)
)
T
The magnetic fields at the mains frequency should
be the values typical of business and hospital
environments.