Chapter 3 - How to...
3.4.8.5 Cable management
Do 
NOT
plugged directly into a wall socket.
To prevent electric shock:
l
l
l
WARNING
l
l
If the mains cable is damaged do not use the system and contact your Optos
representative, see Contact us on page 59.
Ordinary equipment, without protection against ingress of water.
You must regularly inspect the equipment. If there is any damage please contact your
Optos representative, see Contact us on page 59.
CAUTION
Ensure the mains socket is easily accessible.
3.4.9 Troubleshooting
If you have any issues using your device please refer to the following table and instructions.
Additional information and resources are available on the Optos website. Go to optos.com and register to
access the information and resources.
Detailed instructions on using
application.
Note
In this guide the term P200T is used when referring to all product variants, unless otherwise stated.
If you require further assistance, please contact your Optos representative, see Contact us on page 59 .
When contacting Optos you will need to supply your site number, see What information do I need when
contacting Optos? on page 48 .
TABLE 3: List of problems and solutions
Problem
Possible cause
Capture mode is
Failure of the system self-test
disabled.
disables capture.
Password error, image server
Cannot access the
is
NOT
user login list on the
head is
scan head.
the correct image server IP
address.
Page 46 of 65
English
plug the scan head into a power block or powered table. The scan head must be
Only use cables supplied by Optos.
Cables should
be extended or altered.
NOT
Mains supply must be earthed.
The device may only be serviced by Optos trained personnel.
Care must be taken never to touch exposed parts of the device while in physical
contact with a patient.
OptosAdvance
and
Vantage
switched on or scan
configured with
NOT
applications can be found in the help files for each
Solution
Open the diagnostics page and run the self-test
feature.
Check your user name and password, see Why can't
I login? on the facing page.
Check the image server is switched on and running,
see How do I restart the image server? on page 48.
Check the image server network address and ensure
the scan head is on the same network and configured
to connect to the correct address.
Copyright 2017, Optos plc. All rights reserved.
Part Number: G102748/9GCA