With pump alarms:
System operation
Pump is running with alarm
Pump is running with prior
information to alarm
5
Nurse Call quick check protocol
The following protocol provides the user with a quick check guide to allow users to
confirm device behaviour according to these instructions for use and to ensure the
Nurse Call is functional.
CAUTION
It is recommended that this protocol be applied regularly after installation
(especially before use on a new patient, after cleaning or after moving) and
shall be performed by qualified and appropriately trained staff.
Location
Smart Holder COM
serial number
Test description
Visual inspection
Switch pump ON
and generate an
alarm
(eg: start the pump
with no giving set
installed)
The accessory is declared functional ONLY if all tests do CONFORM.
Note: if one or more tests do not conform, please contact your Fresenius Kabi sales
representative.
Operator name
and signature
Ward name:
Expected results
Nurse Call cable, holder and power cable
are not altered or damaged in any way
Check correct customization of the cable
Check power cable and Nurse Call cable
are connected and operational
A visual and audible alarm is displayed on
the pump
The LED is blinking in yellow on the pump
The pump alarm is transmitted on
connected Nurse Call system
Contact
Electrical continuity
C-R
Bed name:
Date: ___ / ___ / ___
C-COM-NC-R
Compliance
Not
Conform
conform