Dimensions – Weight
Holder
Nurse Call cable
Packaging
8.6
Compliance with standards
Safety of
ElectroMedical
Equipments
EMC
Labeling
Note: The full list of applicable standards is available upon request. The device is
protected against leakage current and do not disturb ECG or EEG devices.
9
Cleaning and disinfecting
CAUTION
Do not use the holder and return the device to your biomedical
department in case contamination remains visible on the device after
cleaning.
Please follow the disinfecting and cleaning best practices to limit risks of
holder damage.
No cleaning or disinfection activities should be performed when device
is connected to the mains power supply.
9.1
Prohibited cleaning or disinfection agents
Do not use cleaning or disinfection agents that contain the following substances as
these aggressive agents may damage the plastic parts of the device and cause the
device to malfunction:
trichloroethylene
abrasive detergents
9.2
Precautions
Clean the holder as soon as it becomes contaminated with tube feed or drugs, and at
least once a week. After cleaning, the holder should be left to dry for approximately 5
minutes before being reconnected to the mains.
The holder must be cleaned after each patient usage by a trained nurse or assistant
nurse.
Dimensions (H × W × D)
Weight
~ 0.45 kg
Approx. 132 mm × 118 mm × 46 mm (without
pole clamp)
~ 0.05 kg
Length: approx. 2m
~ 0.20 kg
-
Conform to
Regulation (EU) 2017/745
Conform to
EN/IEC 60601-1
EN/IEC 60601-1-6
EN/IEC 60601-1-11
Conform to
EN/IEC 60601-1-2
Conform to
EN 1041
EN ISO/ISO 15223-1
Protection against moisture:
Holder: IP32 (drip-protected)
Defibrillation-proof type CF
applied part
Class II equipment