3
Holder installation and removal
CAUTION
Give particular attention to the risk of strangulation with cables and sets, and
with the small parts that could be swallowed or inhaled.
3.1
Installation
3.1.1
Using the pole clamp
The holder can be attached universally, vertically and horizontally.
Turn the pole clamp to the suitable position.
3.1.2
Positioning the holder on a rail, pole, bed or wheelchair
Ensure the holder is positioned so that the display is at the suitable height to ensure
good visibility and orientation in the reading direction (the contact pins are at the
bottom).
Fasten pole clamp firmly on the pole or rail to avoid any movement of the pump.
Ensure that the pump is securely attached and firmly positioned.
3.1.3
Positioning the holder on a table
The holder can be placed on a flat and horizontal table
as indicated in the figure.
Ensure the holder is positioned away from table edges to avoid being accidentally
pushed off the table.
3.1.4
Positioning the pump
Slide the pump down until the grey locking
lever locks the position.
3.1.5
Electrical connection and removal
CAUTION
ONLY use the power cable delivered with the device. The power cable and
power socket must remain accessible at all times to allow emergency power
supply disconnection.
Ensure power cable is not damaged.
8
X
Y
Ø
X, Y min = 10 mm
X, Y max = 35 mm
Ø min = 8 mm
Ø max = 40 mm