RS232 serial communication
INFORMATION
RS232 serial communication connector is only used during service
activities with Amika and Ambix nova Partner software and during
communication with PDMS.
All connections and disconnections must be performed by qualified and
trained personnel.
All IT devices (including computers, hubs and switches) inside the
patient area (<1.5m) must comply with IEC/EN 60601-1 (leakage
current).
IT devices connected outside the patient area (>1.5m) must be at least
IEC/EN 60950 compliant.
Do not disconnect communication cables while data is being
transferred.
6.1
Communication with PDMS
INFORMATION
Before connecting the pump to a hospital information system (PDMS),
please contact your IT or biomedical department.
Fresenius Kabi is not responsible of providing serial communication
cable. If need, please contact your PDMS provider for serial
communication cable in accordance with their specifications (DB9,
USB...).
The serial interface will automatically operational when the pump is
installed on the Smart Holder COM and running.
6.2
USB Maintenance cable connection and removal
INFORMATION
USB Maintenance cable is only used to connect the pump with Amika
and Ambix nova Partner software for service activities.
USB Maintenance cable is not provided within the package and must be
ordered separately.
Use ONLY cables recommended by Fresenius Kabi. Refer to per pump
IFU, section Ordering Information.
Before any maintenance operations, the pump must be switched off
and should not be connected to the patient.
For configuration of IT-network and technical specifications, please
refer to the Amika and Ambix nova Partner User Manual (ref. 7125).