Recommended cleaning and disinfection agents
Didecyldimethylammonium chloride (example: Wip'Anios Excel by Anios).
Please contact the appropriate service, responsible for cleaning and disinfection
products in your establishment for further details.
9.4
Cleaning and disinfection guidelines and protocol
INFORMATION
Do not immerse holder in liquids or let liquids enter device's housing.
Holder is resistant to recommended cleaning agents.
9.4.1
Cleaning Instructions
Prerequisites
All cables are unplugged.
The pump is disconnected from the holder.
The air is at room temperature (20 ° C to 25 ° C).
The operator is wearing suitable protective equipment.
Protocol
1.
Place the holder on a cleaned surface or disposable underlay.
2.
During cleaning, do not turn the holder over to avoid liquid leak in.
3.
Use a ready-to-use wipe to remove any major grime.
4.
Thoroughly wipe down all exposed surfaces (housing, pole clamp screw,
connection area, cables etc.) of the holder, from top to bottom.
A minimum cleaning of 1 minute is recommended (allow to remain visibly wet for 1
minute), until all organic matter is dissolved and removed. Do not allow liquids to run,
leak, or drip into the holder housing. Use cotton wool to clean the contact pins.
5.
Using a fresh ready-to-use wipe, thoroughly wipe down all exposed surfaces. A
minimum cleaning of 1 minute is recommended (allow to remain visibly wet for 1
minute), until all organic matter is dissolved and removed.
6.
Wipe down the cables.
7.
Allow the device to dry completely at room temperature.
9.4.2
Disinfection instructions
Prerequisites
The cleaning protocol has been performed.
All cables are unplugged.
The pump is disconnected from the holder.
The air is at room temperature (20 ° C to 25 ° C).
The operator is wearing suitable protective equipment.
Protocol
1.
Place the previously cleaned holder on a cleaned surface or disposable underlay.
2.
During disinfection, do not turn the holder over to avoid liquid leak in.
3.
Use a ready-to-use wipe to wipe down all exposed surfaces of the holder, making
sure to cover all cracks, crevices, and hard-to-reach areas. Do not allow liquids to
run, leak, or drip into the holder housing.
4.
Using a fresh ready-to-use wipe, repeat steps 3. Ensure that the minimum contact
16