To charge the pump battery or to use the pump on the mains power supply:
1.
Connect the power cable to the holder.
2.
Plug the power cable to the mains socket.
3.
To disconnect, unplug power cable from power socket
and remove power cable from holder.
The mains power supply is indicated by a green light on the holder's front panel as well
as on the pump's front panel (keypad).
A beep is emitted by the pump when the power cable is disconnected.
3.1.6
Attaching and Removing the Quick Guide
A quick guide can be easily attached and removed from the pump
holder as indicated on the figure of the right side.
3.2
Removal
3.2.1
Removing the pump from the pump holder
Push the grey locking lever.
Pull the pump up.
3.2.2
Removing the pump holder
4
Nurse Call connection and removal
The Nurse Call connection is functional only if:
the pump is correctly installed on the holder,
the holder is connected to the power supply,
the Nurse Call cable is correctly plugged.
If the Nurse Call is not functional, pump alarm state is not transmitted.
The Nurse Call system availability and technical compliance are the responsibility of the
hospital.
INFORMATION
The Nurse Call cable is delivered with an unterminated side, which will
require customization following specific requirement and installation as
described below.