4 Pre-reprocessing
18
4
Pre-reprocessing
4.1
Initial start-up and longer breaks in use
➢
An adequate normalization phase must be maintained
following transport before initial operation.
➢
Sterilize the ultrasonic handpiece and accessories prior to
startup [→ 37].
➢
Clean and maintain the ultrasonic handpiece after longer
breaks in use.
4.2
Prior to starting the work day
➢
Flush the water paths for 2 minutes.
4.3
Before each patient
1.
Flush the water paths for 30 seconds.
2.
Attach the ultrasonic handpiece [→ 21].
Operating Instructions BACS125 / BACS126
63 83 066 D3275.201.08.03.02