Preparation for Cleaning
• Disassembly is required.
Disassembling Equipment
Completely remove the swivel adaptor
from any components. The swivel adaptor
itself cannot be disassembled any further.
Cleaning – Manual Equipment: Water, Neutral pH Detergent, Soft Bristle Brush, Towels
Method
1. Prepare neutral pH enzymatic detergent solution (e.g. Endozime® AW Triple Plus with APA
(Ruhof), 1:128 ratio) according to detergent manufacturer's instructions using lukewarm
tap water.
2. Prepare equipment for soaking by disassembling removable parts and loosening
connections.
3. Rinse equipment in warm water before placing into bath.
4. Completely soak equipment in water/detergent solution for 30 minutes maximum.
5. Clean thoroughly with a soft nylon bristle brush. NOTE – If possible, use a disposable
brush.
6. Rinse in warm purified water until all visible substances and residual detergent are
removed. NOTE – Make sure to give special attention to hard-to-reach areas.
7. Thoroughly dry equipment with soft clean towels and use medically compressed air if
needed, to dry channels, crevices and lumens.
8. Inspect the equipment to make sure there is no visible organic debris or residue from
cleaning agent.
Repeat process if any soil is detected.
Drying
• Products should be dry at this point. If wetness or excess liquid is detected, dry with a
soft clean towel.
• Medically compressed air can be used if needed.
Optional Automatic Wash / Disinfect
CAUTIONS
• Alkaline and highly acidic detergents and solutions cause damage to the devices.
• Channels and crevices found on these devices require particular attention during
cleaning.
• Pay special attention to the water quality used throughout reprocessing. Hard water can
damage the surface of the equipment. Avoid using hard water. Instead use purified water
unless otherwise specified.
Limitations on Reprocessing
• Repeated processing has negative effects on these devices and is not recommended for
routine use.
• It is important to have Integra NeuroSpecialists perform routine inspections (twice yearly
is recommended).
See contact information.
INSTRUCTIONS
Cleaning - Automated Equipment: Neutral pH Detergent
Method
1. Prepare equipment for cleaning by disassembling removable parts and loosening
connections.
Inspecting for Cleanliness
Inspect the teeth of the starbursts, the area
around the locking screw, and the base of
the swivel adaptor for organic debris.
11
EN – ENGLISH