Integra MAYFIELD A1115 Manual De Instrucciones página 8

Adaptador giratorio modificado

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 58
EN – ENGLISH
Instructions for Use
1. Mount the Base Unit with the Transitional Member onto the operating table.
2. Attach the Modified Swivel Adaptor to the Transitional Member of the Base Unit by
inserting the adaptor's small starburst torque screw into the threaded hole on the
transitional member as shown in Illustration 5:
CAUTION:
Always confirm the starburst teeth mesh properly (as shown in Illustration 3) Failure
to do so may result in damage to the device
3. Mount the headrest or skull clamp on the swivel adapter by inserting the large starburst
Torque Screw into the threaded hole in the skull clamp or headrest. See Illustration 6:
4. The position of the headrest may be adjusted at the swivel adapter and the base unit to
improve patient comfort or accomodate the surgeon's preference.
CAUTION:
Stabilize the patient's head or the headrest equipment until all the connections are
secured against movement.
Turn torque screw clockwise
to secure. Inspect starburst
teeth to confirm correct
engagement.
Illustration 5
Illustration 6
8
Turn torque
screw clockwise
to secure.
Inspect starburst
teeth to confirm
correct engagement.