EN – ENGLISH
2. Rinse equipment in warm water before placing into washer.
3. Load device into the washer and place small parts in container or tray inside the washer
unit in order to avoid losing small components.
NOTE – Load devices carefully into washer in order to avoid collision.
4. Follow the instructions listed below and set washer machine to these exact parameters:
Phase
Time (Min.)
Pre-wash 1
4:00
Enzyme
4:00
Wash
Wash 1
10:00
Rinse 1
0:30
Thermal
2:00
Rinse**
** Optional phase for disinfection of components – minimum water temperature as
indicated or per worker manufacturer specifications for the thermal rinse cycle.
NOTE – Any deviation from this guideline could result in damage to the equipment as well
as improper cleaning results.
Rinse with purified water. Do not perform if parameters cannot be achieved.
5. Remove from washer and dry completely if needed.
6. Inspect equipment to make sure there is no visible organic debris or residue from
cleaning agent.
Repeat process if any visible soil is detected.
The life expectancy of the MAYFIELD products is expected to be as long as 7 years.
Maintenance and Care
To ensure proper function and to extend the life and performance of the equipment, Integra
LifeSciences recommends the following:
Recommended Action
Return the device to the Integra LifeSciences Repairs
department for detailed inspection and servicing.
Request that Integra NeuroSpecialists perform routine
inspections of the device
In the absence of proper care and servicing of the device, negative effects may be seen after
repeated processing over time which may lead to reduced performance.
Contact information: See the Service and Repair section for contact information on how to
return your device for periodic servicing and to request periodic inspections.
See Inspection and/or Service notes section for routine checks to be performed on the device.
Water Temperature Detergent and Concentration
Cold water
Hot water
60.0ºC (140º F)
Hot water
82.2ºC (180º F)
12
N/A
Neutral pH enzymatic (e.g. Endozime®
AW Triple Plus with APA, Ratio 1:128)
Neutral pH detergent
(e.g. Renu-Klenz™ , Steris, Ratio 1:256)
N/A
N/A
Recommended Frequency
Once / year
Twice / year