aIntenanCe tIPS
A. Keep the inner parts of the handpiece as free as possible from
dust and grinding residue.
B. Remove dirt from the outer surface of the handpiece with a
cloth moistened with alcohol. Prevent moisture from entering
the unit.
C. Only use grinding heads with a dental standard size of
2.332 - 2.350 mm (0.091 - 0.093 inches) and avoid using
rusted or damaged grinding heads. Their false usage may
lead to excessive wear and tear in the ball bearings. Adhere
DE
to the safe speed and the safety measures stipulated by the
manufacturer.
EN
D. Do not expose the handpiece to strong knocks as this may
damage the ball bearings. If the handpiece should fall down,
check whether the grinding head is bent. Pay closer attention
FR
to noise and/or heat generation. Send the unit immediately
in for repair should it be damaged.
IT
Caution: Never spray any cleaning agents /
disinfectants onto the handpiece, and onto the handpiece tip
ES
in particular.
PT
PL
E. Cleaning the handpiece
Apart from cleaning the collet chuck, no further maintenance
is required for the spray handpiece. The handpiece must be
cleaned at least once a day depending on its use. Wipe the
handpiece and the hose using a suitable disinfectant wipe.
Clean the spray nozzle with a small brush without leaving
any moisture.
Kaitstud / Secured by https://mereneid.eu/
F. Instructions for cleaning the chuck:
While pushing the cutter shafts in the chuck it is possible that
the dirt on the cutter dissolves and falls in the chuck. Should it
occur at certain time intervals where dust accumulates so that
cutter shaft is no longer pushed at the intended depth and
thus the full clamping force is not guaranteed, so the chuck is
cleaned as per the following information:
• The delivery contains a special spiral driller with handle to
perform a simple cleaning.