lasts about 12 minutes. It includes the additional differentiated testing of the internal tanks and
should be performed on a weekly basis.
What to do
How to do it
1. Call up the Main
Press or turn the select knob
Menu
to call up the Main Menu.
2. Set up the self-
Turn the select knob to the
test
"Self-Test" setting.
3. Confirming the
Press the "Execute" button.
self-test
The "Self-Test" submenu is
called up. The last test per-
formed is shown with a frame
around it.
4. Select the Quick
Turn the select knob to the
Test
"Quick Test" setting.
5. Confirm the
Press the "Start/Execute" but-
Quick Test
ton.
The start menu is called up.
6. Seal the system Press the supplied cap firmly
into the adapter (for the mul-
tichannel adapter, seal all four
channels).
7. Start the Quick
Press the "Start/Execute" but-
Test
ton.
The five components – sensor,
pump, valve, filter, and mea-
surement chamber – are
tested in succession.
PLT unit 997960
Button to use
4 The first steps
Display
Main Menu
Pipette Type
Self Test
Switch OFF
Make settings
select
back
Main Menu
Pipette Type
Self Test
Switch OFF
Make settings
select
back
Self Test
Quick Test
Full Test
select
back
Self Test
Quick test
Full test
select
back
Self Test
Please insert
blind plug
select
back
Self Test
-0.10 hPa 1:sensor
back
Operating manual