5 Performing the test
5.4 Testing multichannel pipettes
Example: 200 μL 8-channel air-displacement pipette, dynamic test
Before performing the test, set the pipette to the nominal volume. Before testing, mount new unused
disposable tips and insert four channels with mounted tips into the previously selected and attached
Adapter 3. Select an adapter; see Test preparation, p. 52.
What to do
1. Call up the Main
Menu
2. Set up the self-
test
3. Confirm the
pipette type
4. Select the Quick
Test
5. Confirm the sin-
gle-channel
pipette
6. Select the
pipette volume
60
Operating manual
How to do it
Press or turn the select knob
to call up the Main Menu.
Turn the select knob to the
"Pipette Type" setting.
Press the "Execute" button.
The "Channels" submenu is
called up.
Turn the select knob to the
"Multichannel" setting.
Press the "Execute" button.
The "Nominal Volume" menu
is called up.
Turn the select Select to the
"200 μl" setting.
Button to use
Display
Main Menu
Pipette Type
Self Test
Switch OFF
Make settings
select
back
Main Menu
Pipette Type
Self Test
Switch OFF
Make settings
select
back
Channels
Single Channel
Multichannel
select
back
Channels
Single Channel
Multichannel
select
back
Nominal Volume
25 µl
50 µl
100 µl
200 µl
select
back
Nominal Volume
25 µl
50 µl
100 µl
200 µl
select
back