Errors occurring during self-test
Self Test
Error 1xxx
See instruction manual
Error code
1000
1010
1020
1040
1050
1030
1060
1070
1080
1090
1020
1150
1160
1170
1200
11 Product markings
Symbol or Number
www.brand.de/ip
XXNXXXXX
PLT unit 997960
back
Cause
Pressure sensor defective
Pump running continuously Send device in for repair.
Pumping out taking too
long
Valve defective
Aeration through aeration
channel taking too long
Device leaking
Aeration through adapter
taking too long
Electronics fault
Meaning
With this mark, we confirm that the product complies with the require-
ments set out in the EC Directives and has undergone testing according to
the specified test procedures.
UKCA: United Kingdom Conformity Assessed
With this mark, we confirm that the product complies with the require-
ments set out in the UK Designated Standards.
Patent information
Serial number
Remedy
Send device in for repair.
Check whether cap is securely inserted and
adapter is screwed in tightly. If so, then pump is
defective. Send device in for repair.
Send device in for repair.
Send device in for repair.
Check cap, adapter seating, and adapter filter,
and replace if necessary. Repeat self-test. If er-
rors cannot be resolved, send the device in for
repair.
Remove cap if not already done.
If adapter filter is contaminated, replace filter.
Send device in for repair.