Switch Contacts; Wiring Scheme - WIKA GDM-RC-100 Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

5. Commissioning, operation
5.2.3 Connection terminals and wire preparation
The connection terminals are suitable for the following types of single conductor or
cross-sections:
EN
Connection terminals and wire preparation
Types of single
conductors
Wire length
Maximum length of
the bare wire end
Number of wires /
Cross-sections
Recommended
tightening torque

5.2.4 Switch contacts

The position of the switch points and the switching functions are indicated on the
product label.
The assignment of the individual switching functions is marked on a sticker affixed to
the mating connector at the terminal:

Wiring scheme

First digit
Second digit
14
Connection terminals of
the socket terminal strip
Solid wire core
Flexible strand
Flexible strand with end splice
≥ 90 mm
Max. 5.5 mm
1 x 0.5 mm² to 1 x 2.5 mm²
0.5 Nm
1x – Switch contact S1
2x – Switch contact S2
3x – Switch contact S3
4x – Switch contact S4
x1 – Common
x2 – Normally closed
x3 – Normally open
Connection terminals of
the protective conductor
Solid wire core
Flexible strand with end splice
≥ 120 mm
Max. 8 mm
1.2 Nm
WIKA operating instructions, model GDM-RC-100

Publicidad

Tabla de contenido