Descargar Imprimir esta página

HEBU medical HF 9514 Instrucciones Para El Uso página 54

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

EN
Argon gas flow setting
The gas flow can be set at an interval of 0.1 L/min to 6.0 L/min in steps of 0.1 L/min.
Different gas flows can be set for each of the functions CUT and COAG.
To set the gas flow, touch the
relevant display - CUT or COAG.
This will open the menu for setting
the gas flow. Use the arrow keys to
set the required output.
To save and return to the main menu, select the
"Exit" button.
8.9 Saving and accessing the program
Touching the "Program" field causes a selection of saved
programs to appear.
To select the required program touch the relevant line. The
program you are located in is highlighted in grey.
In the main menu, the number of the accessed program and
the name are displayed
To change the program name, touch the field with the name.
In our example, this is: DR MULLER. A window opens with a keypad. The maximum symbol length
is 15 characters.
To save the name, press "Enter", and to quit without saving, press "Esc".
Every change of the settings is automatically saved in the
program you are currently in.
In order to save individual settings in a new program, initially
select a free storage location. Settings and the name of the
program can then be changed at will.
54
HBS x-touch 300 argon
DR. MULLER
1

Publicidad

loading

Este manual también es adecuado para:

Hbs x-touch 300 argon