HBS x-touch 300 argon
9 Guideline value for power settings
When performing the power settings, the user should generally
take into account the cutting speed, geometry of the electrode
Monopolar cutting:
Smooth cuts with thin needle electrode
Cutting with knife electrode, lancet or wire
loop
Cutting in wet environments (TUR)
Monopolar coagulation, spray coagulation:
Coagulation of large surfaces
Coagulation of small surfaces
Argon coagulation:
With flexible tube
With rigid electrode
Bipolar application:
Coagulation
Blend Cut
Cutting in water environment (TUR)
The recommended settings serve as guideline values only and
and properties of the tissue.
must be adjusted to the relevant situation.
In the event of unfavourable conditions, these values may
Recommended settings
5 - 40 W
50 - 175 W
80 - 150 W
Recommended settings
60 W
40 W
Recommended settings
20 – 40 W 1,5-2,5 L/min
25-50 W 2,5-4,5 L/min
Recommended settings
30 - 60 W
40 - 60 W
50 - 75 W
deviate!
EN