DJO Global cefar easy Manual Del Usuario página 180

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 63
(9.397 pt)
Immunity Test
Electrostatic
discharge (ESD) IEC
61000-4-2
Electrical fast
transient/burst IEC
61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
FI
Power frequency
(50/60Hz)
magnetic field
IEC 61000-4-8
NOTE: U T is the AC mains voltage prior to application of the test level.
178
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
IEC 60601
Test Level
±6kV contact
±8kV air
±2kV for power supply lines
±1kV for input/output lines
+ 1kV differential mode (line to line)
+ 2kV common mode (line to ground)
<5% U T (>95% dip in U T ) for 0,5 cycle
40% U T (60% dip in U T ) for 5 cycles
70% U T (30% dip in U T ) for 25 cycles
<5% U T (>95% dip in U T ) for 5 sec
3 A/m
Electromagnetic Environment -
Compliance Level
Risk assessment on the CEFAR EASY
Electrotherapy System indicates
±6kV contact
the compliance levels claimed are
±8kV air
acceptable when ESD-precautionary
measures are taken.
Not Applicable -
Battery powered
Mains power quality should be that
of a typical commercial or hospital
Not Applicable
environment.
- signal lines less
then 3 meters
Mains power quality should be that
Not Applicable -
of a typical commercial or hospital
Battery powered
environment.
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the
CEFAR EASY Electrotherapy System
Not Applicable -
requires continued operation during
Battery powered
power mains interruptions, it is
recommended that the CEFAR EASY
Electrotherapy System be powered
from an uninterrupted power supply.
3 A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital environment.
Guidance

Publicidad

Tabla de contenido
loading

Tabla de contenido