STARTER 3100 Bench pH Meter
-
Slope: 99%
Offset: 5mV
ATC/MTC: MTC
The print-out for a 3-point calibration is:
Buffer1: 4.01pH
mV1: 178mV
Temp.1: 25.0 ° C
Buffer2: 7.00pH
mV2: 0mV
Temp.2: 25.0 ° C
Buffer3: 9.21pH
mV3: -130mV
Temp.3: 25.0 ° C
Slope: 100%
Offset: 0mV
ATC/MTC: ATC
Printing from memory: When scrolling through the memory you can print the entry that is currently
viewed by pressing and holding (3 seconds) the
. The printout format is as follows:
Recall 01:
End Point, Value, Temp., ATC/MTC
Manual EP, 4.01pH, 35.6 ° C, ATC