STARTER 3100 Bench pH Meter
Press button
result is being displayed.
Press
4.4.3
Clearing the memory
Pressing
Press
There are now 2 options:
Press
Press
4.5
Printing
If the printer is connected to the STARTER 3100(e.g. SF-F40A, green light means connection
success; printer baud rate 9600bps; 8 data bit; none parity bit; 1 stop bit), a print-out is
automatically generated after each end pointed measurement or calibration.
The format for the print-out following a pH measurement is:
End Point, Value, Temp., ATC/MTC
Auto EP, 4.01pH, 25.0 ° C, MTC
The details for the second line are:
Auto EP, 4.01pH, 25.0 ° C , MTC
|
|
|
|
|
|------------------------- pH value
|---------------------------------------- Auto End Point
The format for the print-out following a mV measurement is:
End Point, Value, Temp., ATC/MTC
Manual EP, 182mV, 23.2 ° C, ATC
The print-out in case of an error message is:
End Point, Value, Temp., ATC/MTC
Error3
The print-out for a 2-point calibration is:
Buffer1: 4.01pH
mV1: 178mV
Temp.1: 25.0 ° C
Buffer2: 7.00pH
mV2: 3mV
Temp.2: 25.0 ° C
or
to scroll through the stored results. R01 to R99 indicates which
to exit.
to scroll through the stored results until "
or
, CLr blinks;
to confirm the deletion of all the stored data.
to return to the measurement mode without deleting the memory.
|
|---- Manual Temperature Compensation
|---------------- Temperature value and unit º C
EN-13