Table 1: Validated Washer Parameters for the AR-1650-02 & AR-1650-03.
Phase
Pre-Wash 1
Wash 1
Rinse 1
Rinse 2
Thermal Disinfection
Drying
Steam Sterilization for AR-1650-02 & AR-1650-03
Sterilizer Type:
Preconditioning Pulses:
Minimum Temperature:
Full Cycle Time:
Minimum Dry Time:
Cool-Down Time:
Sterilizer Type:
Minimum Temperature:
Full Cycle Time:
Minimum Dry Time:
Cool-Down Time:
Manufactured For:
Arthrex, Inc.
1370 Creekside Blvd
Naples, FL 34108, USA
1-(800) 934-4404
Recirculation time
(minutes)
3:00
10:00
3:00
3:00
5:00
6:00
Prevacuum
4
132°C
4 minutes
30 minutes
30 Minutes outside of chamber on a wire rack
Gravity
132°C
15 Minutes
50 Minutes
30 minutes outside of chamber on a wire rack
Arthrex GmbH
Allen Medical Systems
Erwin-Hielscher-Strasse 9
100 Discovery Way
Acton, MA 01720
81249 München, Germany
(978) 266-5401
+49 89 909005-0
Temperature
Cold tap water
65.5°C
Neodisher® Mediclean Forte
50°C
50°C
90°C
115°C
Hill Rom SAS
B.P. 14 Z.I. du Talhouet
56330 Pluvigner, France
Tel. +33 (0) 2 97 50 92 12
Fax. +33 (0) 2 97 50 92 00
Detergent type and
Concentration
N/A
¼ oz/gallon
N/A
N/A
N/A
N/A
D-720771-A1
08 FEB 2016