QC MANAGER SETUP
(Page 1 of 2)
Location
[CVOR]
User ID Setup
Clear ID Interval
[20 minutes]
The [Location] variable function key is used to select the location of the ACT Plus
instrument. Location options include the following: CVOR (cardiovascular
operating room), Cath Lab (Cardiac Catheterization Lab), ECMO (Extracorporeal
Membrane Oxygenation), Dialysis, ICU/CCU (Intensive Care/Coronary Care
Units), Lab, and Other.
The [User ID Setup] variable function key displays the screens for the following
functions: adding and deleting User IDs, turning ON or OFF User Lockout,
changing the QC Manager Password, downloading User IDs to a floppy disk or
USB storage device, and uploading User IDs from a floppy disk or USB storage
device to the ACT Plus
The [Clear ID Interval] variable function key is used to select the time interval
(none, 5 minutes, 10 minutes, 20 minutes, 60 minutes, or 120 minutes) that will
elapse before the Patient and User IDs are cleared. Once cleared, the PID and UID
will need to be reentered prior to testing.
The [QC Lockout] variable function key is used to set QC Lockout to 1 of the
following options: OFF, Warning, or ON.
The [QC Interval Setup] variable function key is used to set the required interval
for performing liquid and electronic QC tests. The options for liquid are the
following: None, 8 hours, or 7 days. The options for ACTtrac are None or 8 hours.
The [Cartridge Enable Setup] variable function key is used to enable or disable
the various cartridge types that may be run on the ACT Plus
The [Download Settings] variable function key is used to download all instrument
parameters and settings to a floppy disk or USB storage device.
The [Perm Record] variable function key designates the official mode of tracking
the results that have been downloaded. The options are None, Floppy, USB, or
Electronic.
The [Instrument Upgrade] variable function key is used to install upgrades to the
®
ACT Plus
software.
®
ACT Plus
Operator's Manual
Figure 6. QC Manager Setup Screen
®
user database.
--->> QC Manager <<---
QC MANAGER SETUP
(Page 2 of 2)
Cartridge
Enable Setup
Download
Settings
Go to
Page 1 of 2
®
Perm Record
[USB]
Instrument
Upgrade
Exit to Quality
Control Menu
instrument.