Guidance and manufacturer's declaration - electromagnetic immunity
®
The ACT Plus
is intended for use in the electromagnetic environment specified below.
The customer or the user of the ACT Plus
environment.
Immunity test
IEC 60601
test level
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80
MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2.5
GHz
Note: At 80 MHz and 800 MHz, the higher frequency range applies.
Note: These guidelines may not apply in all situations. Electromagnetic propagation
is affected by absorption and reflection from structures, objects and people.
®
ACT Plus
Operator's Manual
IEC 60601-1-2: 2001 Table 204
®
should assure that it is used in such an
Compliance
Level
3 Vrms
3 V/m
Electromagnetic environment -
guidance
Portable and mobile RF
communications equipment should be
used no closer to any part of the ACT
®
Plus
, including cables, than the
recommended separation distance
calculated from the equation applicable
to the frequency of the transmitter.
Recommended separation distance
√
d = 1.2
P
√
d = 1.2
P 80 MHz to 800 MHz
√
d = 2.3
P 800 MHz to 2.5 GHz
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation distance in
meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey,
less than the compliance level in each
b
frequency range.
Interference may occur in the vicinity of
equipment marked with the following
symbol:
a
should be
English 59