Expiration date is not valid.
OR
Invalid expiration date.
Test will overwrite saved test history.
Invalid password.
®
Run ACTtrac
from QC Menu.
No cartridge lot number active.
Cartridge not recognized.
OR
There are no active cartridges to edit.
OR
There are no active cartridges.
No control lot number active.
There are no other active cartridges.
No test results available.
Patient ID required.
Preventive maintenance due.
Value Out of Range. Enter 35 – 39.
There are no intervals set.
QC Due: HH:MM.
QC not performed.
®
ACT Plus
Operator's Manual
The expiration date is not in a valid format.
The proper format for the expiration date for
a cartridge or control lot is yy-mm-dd.
The current test will overwrite test results in
the permanent record. Transmit all unsent
test records to the output source.
The password that was entered is not
correct. Enter the correct password.
The user is attempting to run an ACTtrac
test from the Main Menu. Go to the QC Menu
screen to run ACTtrac
The user is attempting to run a test without
an active cartridge lot number. Enter the
cartridge lot number for this test type.
All cartridges have been disabled. Have the
QC Manager enable the appropriate tests.
The user is attempting to run a control test
without an active control lot number. Enter a
valid lot number for the control.
Only the selected test is enabled. Have the
QC Manager enable tests.
No test results are available to view on the
View Current Test Results screen.
Patient ID is required to run a patient test.
Enter a Patient ID in the Enter ID screen.
Preventive maintenance is due. Call
Medtronic Instrument Service to schedule a
preventive maintenance check.
The Actuator Heat Block temperature cannot
be adjusted out of the 35-to-39 range. If the
Actuator Heat Block cannot be adjusted
within the 35-to-39 range, call Medtronic
Instrument Service.
The QC Manager has not set the intervals
for liquid or electronic controls. Have the QC
Manager set QC intervals in the QC
Manager screen.
The QC for this cartridge will run out at the
prescribed time (hh = hour; mm = minute).
The QC test for this cartridge has not been
performed or has failed. Perform QC tests
for this cartridge.
®
®
tests.
English 51