Descargar Imprimir esta página

Ecleris PROCAM HD BASIC Manual De Usuario página 10

Sistema cámara y consola

Publicidad

Idiomas disponibles

Idiomas disponibles

The following procedures are recommended to conserve PROCAM HD, in optimum working
conditions:
Before the maintenance of the PROCAM HD, disconnect the power cable from the device
and from the electric power supply.
Periodic cleaning of all external components.
Check connections.
Annual technical verification of the equipment.
For any type of maintenance work or revision contact your local distributor or nearest ECLERIS
office.
6.1 CLEANING
Before cleaning the PROCAM HD, disconnect the power cable from the device and from
the electric power supply.
6.1.1 EXTERNAL SURFACE
To clean the outer surfaces of the device only use cotton or cloth dampened in mild detergent or
soap. Only use adequate cleaning products prepared for electrical equipment following the
products instruction.
While cleaning, be careful not to let water or humidity get into the equipment.
Never use flammable products, especially alcohol.
Do not turn the equipment on until one (1) hour after cleaning process.
We recommend cleaning the external surface of the equipment every 15 days.
Do not sterilize any parts of the device.
Do no immerse the device or its accessories.
The cleaning, disinfection and sterilization of accessories not provided by ECLERIS must
be performed following the instruction provided by their manufacturer.
6.2 ELECTRICAL CONNECTIONS
Permanently check the condition of all the electrical components.
Check that there are no:
Cables in bad conditions; cut or torn.
Bad connection of the cable with the socket.
Rusty pins in the connectors.
Disconnected equipotential ground connection (in places with equipotential ground).
6.3 CHANGING THE FUSE
To avoid a waste of time for unavailability, we recommend having spare fuses on hand.
1. Disconnect the power cable from the electric current and from the device.
INMANUUML023 USER M./M.USO PROCAM HD BASIC V.2022.10 ENG/ESP US
6. MAINTENANCE
USER MANUAL – PROCAM HD II
Página 10 de 30

Publicidad

loading