Guidance and Manufacturer's Declaration – Electromagnetic Emissions
The CARINA
Digital Intraoral Sensor, used with a compliant computer, is
TM
intended for use in the electromagnetic environment specified below.
The customer or the user of the CARINA
is used in such an environment.
Emissions Test
Compliance
RF emissions
Group 1
CISPR 11
RF emissions
Class B
CISPR 11
Harmonic emissions
Class A (*)
IEC 61000-3-2
Voltage fluctuations/
Complies (*)
flicker emissions IEC
61000-3-3
(*) Computer used with the CARINA
sensor must meet this rating.
TM
10 |
henryscheinequipmentcatalog.com/carina
Digital Intraoral Sensor should assure that it
TM
Electromagnetic Environment – Guidance
The sensor uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in
nearby electronic equipment.
The sensor is suitable for use in all establishments
including domestic establishments and those
directly connected to the public low-voltage power
supply network that supplies buildings used for
domestic purposes.
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
The CARINA
TM
use in the electromagnetic environment specified below. The customer or the user of the
CARINA
Digital Intraoral Sensor should assure that it is used in such an environment.
TM
Immunity
Test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
Power frequency
(50/60Hz)
magnetic field
IEC 61000-4-8
NOTE 1 UT is the mains voltage prior to application of the test level.
(*) Computer used with the CARINA
English
English
Digital Intraoral Sensor, used with a compliant computer, is intended for
IEC 60601
Compliance
Test Level
Level
+/- 8 kV contact
Complies
+/- 2 kV, +/- 4 kV, +/- 8 kV,
+/- 15 kV air
+/- 2 kV for power supply lines
Complies
(*)
+/- 1 kV for input/output lines
+/- 1 kV differential mode
Complies
(*)
+/- 2 common mode
0% UT; 0.5 cycle at 0o, 45o, 90o, 135o,
Complies
180o, 225o, 270o, and 315o 0% UT;
(*)
1 cycle
70% UT; 25/30 cycles for 50 Hz and
60 Hz, respectively Single phase:
at 0o
0% UT; 250/300 cycle for 50 Hz and
60 Hz respectively
Single phase: at 0o
VDI specs per 3rd edition:
<5% UT (>95% dip in UT) for 0,5 cycle
40% UT (60% dip in UT) for 5 cycles
70% UT (30% dip in UT) for 25 cycles
<5% UT (>95% dip in UT) for 5 s
30 A/m, 50 Hz or 60 Hz
Complies
sensor must meet this rating.
TM
henryscheinequipmentcatalog.com/carina
Electromagnetic
Environment –
Guidance
Floors should be
wood, concrete, or
ceramic tile. If floors are
covered with synthetic
material, the relative
humidity should be at
least 30%.
Mains power quality
should be that of a
typical commercial or
hospital environment.
Mains power quality
should be that of a
typical commercial or
hospital environment.
If the user of the sensor
requires continued
operation during power
mains interruptions, it
is recommended that
the sensor be power
from an uninterruptible
power supply or a
battery.
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in
atypical commercial
or hospital
environment.
| 11