APPENDIX B — SPECIFICATIONS AND STANDARDS
Sensor Architecture
Indirect converting dental intraoral CMOS x-ray sensor
Image Pixel Size: 19.5µm
Image Resolution: 20+ visible lp/mm
Maximum Gray Levels: 16,384 (14 bit)
X-Ray Parameters
Sensor can be used with dental x-ray generators in the range of 60 to 70
kV; at minimal 40 µGy incident dose
Electrical Rating
DC 5V, 0.5W
USB 2.0 compliant
Connection to PC
Cable Length
1.9m
Degree of Protection
Type BF
Against Shock
Applied part
Mode of Operation
Continuous
Method of Sterilization
Sensor not suitable for sterilization or autoclave
Type of Protection
Class II
Against Electrical Shock
Suitability for Use in an
Not intended for use in an Oxygen Rich Environment
OXYGEN RICH
ENVIRONMENT
Protection Against
IPX8 (Protected against the effects of continuous immersion in water at
Harmful Ingress of Water
1.1 meters for up to 35 minutes)
Sensor Technology
CMOS
External Dimensions
(mm)
CARINA
38.83 x 29.63 x 13.01
TM
Environmental
Humidity
Conditions
Usage
30% to 75%
CARINA
TM
units are not suitable
to be operated in oxygen rich
and/or explosive environments
Transportation and Storage
10% to 95%
Transport in the supplied pro-
tective pack
Packaging Labels
14 |
henryscheinequipmentcatalog.com/carina
Active Imaging Area
Active Image Area
(mm)
(px)
32.99 x 25.82
1692 x 1324
Air
Ambient
Pressure
Temperature
700 hPa to 1060
5° to 30°C
hPa in altitudes
up to 3000
meters
-40° to 70°C
English
APPENDIX C — PRODUCT SYMBOLS
Symbol
Title of the
Refer-
Standard
Symbol
ence
Contain-
Number
ing the
Symbol
Type BF
5333
IEC 60417
applied part
Direct
5031
IEC 60417
Current
General
W001
ISO 7010
warning sign
Radioactivity
W003
ISO 7010
and ionizing
radiation
Caution
0434B
ISO 7000
Do not reuse
1051
ISO 7000
Marking of
N/A
BS EN
electri-
50419
cal and
electronic
equipment
Manufac-
3082
ISO 7000
turer
English
henryscheinequipmentcatalog.com/carina
Fuction/ Description
Manufacturer
per Standard
Interpretation
To identify a type BF
To identify a type BF
applied part comply-
applied part complying
ing with IEC 60601-1
with IEC 60601-1
To indicate on the
Direct Current
rating plate that the
equipment is suitable
for direct current only;
to identify relevant
terminals
To signify a general
Warning
warning
To warn of radioactive
Radiation Warning
materials or ionizing
radiation
To indicate that
Please refer to the
caution is necessary
written instructions in
when operating the
this manual
device or control close
to where the symbol is
placed, or to indicate
that the current situ-
ation needs operator
awareness or oper-
ator action in order
to avoid undesirable
consequences.
To indicate that the
Please contact your
item is for single use
supplier for further
only and must not
information about the
be used more than
product disposal at the
once, for example, on
end of the product's
packages of medical
lifetime
disposables
This symbol on the
Please contact your
products and/or
supplier for further
accompanying docu-
information about the
ments means that the
product disposal at the
used electrical and
end of the product's
electronic products
lifetime
should not be mixed
with general house-
hold waste
To identify the manu-
Manufacturer
facturer of a product.
This symbol shall be
used filled in all appli-
cations to differentiate
it from ISO 7000-2497