8 Maintenance and service
48
If the values measured during one of the three calibration checks lie outside
of the permissible range, please proceed according to the following flow chart:
Start
One of the three
measured values is
outside the permissible
range.
Has the optical fiber
been prepared
properly?
YES
Is the optical fiber
aimed exactly at the
sensor range?
he power meter
is switched on and
set so that a power of up to
10 W can be correctly
measured at 970 nm ± 10 nm.
YES
Does the error
persist when repeating
the measurement?
NO
OK
Cut the end of the
NO
optical fiber so that an
exact circle becomes
visible.
Aim the optical fiber at the
measuring sensor from a
NO
distance of at least 2 cm,
making sure that the red
circle is located inside the
sensor range.
Set the power meter to a
measuring range of max.
NO
10 W (upper end of scale)
and a wavelength of
970 nm +/– 15 nm.
Sirona Dental
YES
Systems LLC,
Dental Sales or
authorized
service center.
60 56 092 D 3485
D 3485.201.01.10.02 03.2008