Figure 5:
Distal part of handpiece
Figure 6:
Distal part with finger
switch
Figure 7:
Finger switch mounted on
handpiece
Figure 8:
Optical fiber and guide
rings
Figure 9:
Connection of the finger
switch
60 56 092 D 3485
D 3485.201.01.10.02 03.2008
4.7
Connection of the finger switch
Unscrew the distal part of the handpiece.
Fasten the finger switch.
Screw the distal part of the handpiece back onto the remainder of the
handpiece.
Run the optical fiber through the guide rings on the finger switch supply
cable.
Plug the connector of the finger switch cable into the socket of the Y
connector or into the "Interlock/Switch" socket.
The treatment room must be protected by suitable measures
(in compliance with IEC 60825-1), e.g. by closing the doors.