10 Manufacturer's declaration on electromagnetic compatibility
RF immunity
The SIROLaser is intended for use in the electromagnetic environment specified below. The user should ensure that it is
used in such an environment.
Immunity test
Conducted RF interference
EN 61000-4-6
Radiated RF interference
EN 61000-4-3
According to a study on electromagnetic systems, the field strengths of permanently installed RF transmit-
ters may lie within any frequency range below the problematic range.
Interference may occur in the area surrounding devices marked by the adjacent symbol.
54
EN 60601-1-2
Compliance level
test level
3 Vrms from 150
3 Vrms from 150 kHz
kHz to 80 MHz
to 80 MHz
3 Vrms from 80
3 Vrms from 80 MHz
MHz to 2.5 GHz
to 2.5 GHz
Electromagnetic environment
- Explanation
Portable and mobile RF communication
equipment should be located at least as
far away from any component of the
laser unit, including its power cord, as is
indicated by the equation correspond-
ing to the relevant transmitter fre-
quency.
d = 1.2 Þ÷P from 80 MHz to 150 MHz
d = 1.2 Þ÷P from 80 MHz to 800 MHz
d = 2.3 Þ÷P from 800 MHz to 2.5 GHz
where "P" is the maximum output power
rating of the transmitter in watts accord-
ing to the transmitter manufacturer and
"d" is the recommended separation dis-
tance in meters.
60 56 092 D 3485
D 3485.201.01.10.02 03.2008