Temperature Dry Well Calibrators/Micro Calibration Baths
CTD9100/CTB9100
6.3.1 Switching off the automatic control
For certain tasks it can be advantageous to switch
off the control, e.g. to carry out settings at the
calibrator/micro calibration bath.
Press the P key when in calibration mode for
approx 5 sec., the main menu opens.
OPEr appears on the upper PV display
LED SET flashes on the lower SV display.
Fig. 20: Main menu display
Press the 5 or  key until OFF appears.
Fig. 21: Menu control OFF
Press the P key to confirm.
An alternating display of the current reference
temperature and OFF appears on the upper PV
display
The current set temperature appears on the
lower SV display.
Fig. 22: Control OFF setting display
The control has now been
switched off and the reference
temperature will constantly drop
without being regulated.
WIKA Operating Instructions Temperature Dry Well Calibrators/Micro Calibration Baths
6.3.2 Switching on the automatic control
The control is switched off if the following display
appears:
An alternating display of the current reference
temperature and OFF appears on the upper PV
display.
The current set temperature appears on the lower
SV display.
Fig. 23: Control OFF setting display
Switch the control back on by:
Pressing the P key for approx. 5 sec., the main
menu opens.
rEG appears on the upper PV display
LED SET flashes on the lower SV display.
Fig. 24: rEG display
Press the P key to confirm switching on the
control.
The control has been reactivated.
The calibrator/micro calibration
bath is in calibration mode and the
set temperature is targeted.