WIKA CTD9100-165 Manual De Instrucciones página 7

Calibradores de temperatura/microbanos de calibración

Publicidad

Idiomas disponibles

Idiomas disponibles

Temperature Dry Well Calibrators/Micro Calibration Baths
CTD9100/CTB9100
2.3 Safety instructions for the application of
calibration liquids
Calibration liquid water:
Only use distilled water, otherwise excessive
limescale and soiling will build up in the
calibrator tank.
Calibration liquid silicone oil:
Only use the silicone oil recommended in these
operating instructions.
Always read the safety data sheet supplied with
the silicone oil before using it.
Always ensure adequate ventilation when
working with silicone oil, since hazardous
substances can be released.
Prevent silicone oil from coming into contact with
your eyes.
Since silicone oil is hygroscopic, always use the
transport lid to close the calibration bath after
use.
The transport lid is equipped with a safety
valve. If the micro calibration bath is closed
when warm, impermissible pressure can build
up. In order to prevent excess pressure which
can destroy the liquid bath, the safety valve is
activated once the pressure reaches approx. 2.5
bar. This can result in hot steam being released.
If problems or questions arise, please contact your
supplier or the manufacturer directly:
Alexander Wiegand GmbH & Co. KG
Alexander-Wiegand-Straße, 63911 Klingenberg,
Germany
Tel:
+49 9372 132 9986,
Fax:
+49 09372 132 217
Email: testequip@wika.de
www.wika.de
WIKA Operating Instructions Temperature Dry Well Calibrators/Micro Calibration Baths
Risk of severe burns!
Prior to transport or contact with
the metal block/liquid bath ensure
that it has cooled down sufficiently,
otherwise there is a risk of severe
burns caused by the metal block/
liquid bath and the test specimen.
GB
7