9. Cleaning and Maintenance
GB
Allow the calibrator/micro calibration bath to cool
down as described in chapter 8.
Switch the calibrator/micro calibration bath off
and disconnect the mains plug.
Cleaning calibrators with sleeves:
A small amount of brass dust is created when
operating calibrators with sleeves, this can cause
the metal block and sleeve to jam. To prevent
this, remove the sleeves from the heating block at
regular intervals and if the calibrator is not going to
be operated for a longer period. Flush the heating
block bore with compressed air and clean the bore
and sleeve with a dry cloth.
Cleaning the fan grille:
Each calibrator is fitted with a small meshed air
grille via which cooling air enters the calibrator.
10. Warranty and Repairs
The calibrator/micro calibration bath is under
guarantee for 12 months as from the date of
delivery for construction errors or material defects.
The guarantee is limited to repairs or replacing the
calibrator/micro calibration bath.
Warranty shall not apply if the calibrator/micro
calibration bath is opened and unauthorized
repair work is carried out or if the calibrator/micro
calibration bath is not used for its intended purpose
or installed incorrectly.
11. Recalibrating
The calibrator/micro calibration bath is adjusted and
tested with measuring equipment in accordance
with recognized national standards prior to delivery.
The calibrator/micro calibration bath should,
depending on the application situation, be
inspected at appropriate intervals on the basis of
DIN ISO 10 012. We recommend you to return the
26
WIKA Operating Instructions Temperature Dry Well Calibrators/Micro Calibration Baths
Clean the grille at regular intervals (vacuuming or
brushing) depending on the level of air pollution.
Cleaning the micro calibration bath:
Drain as much of the silicone oil as possible.
Subsequently remove the sensor cage from the tank
and clean the cage, magnetic stirrer and the tank
with water and plenty of washing-up liquid. Allow
everything to dry completely.
If you are using distilled water, remove the
calibration liquid and allow the sensor cage,
magnetic stirrer and tank to dry completely.
External cleaning:
Clean the outside of the calibrator/micro calibration
bath with a damp cloth and some water or with a
solvent-free mild cleaning agent.
If the calibrator/micro calibration bath malfunctions
during or after the warranty period, always contact
the WIKA Test & Calibration Department before
sending the calibrator/micro calibration bath for
repairs.
The defective calibrator/micro calibration bath incl.
details of the occurred fault can be sent freight paid
to WIKA, unless other agreements have been made.
calibrator/micro calibration bath to WIKA at intervals
of max. 12 months or approx. 500 operating hours
for recalibration and readjustment.
Recalibration is based on the directive DKD R5-4
of the German Calibration Service. The measures
described here are applied and considered during
recalibration.