Temperature Dry Well Calibrators/Micro Calibration Baths
CTD9100/CTB9100
6.3.5 Setting and saving fixed set temperatures
In order to save set temperatures in the calibrator/
micro calibration bath, the respective set value
memory has to be opened.
Press the P key for approx 5 sec. when in
calibration mode, the main menu opens.
OPEr appears on the upper PV display.
LED SET flashes on the lower SV display.
Fig. 30: Operator menu OPEr
Press the P key again, the group level opens.
OPEr appears on the upper PV display.
'SP appears on the lower SV display and LED
SET flashes.
Fig. 31: Group SP
Press the P key again, the parameter level
opens.
'SP appears on the upper PV display.
The set value memory SP1 and LED SET flash
on the lower SV display.
Fig. 32: Parameter for the set memory SP1
WIKA Operating Instructions Temperature Dry Well Calibrators/Micro Calibration Baths
Use the 5 or  key to select one of the four set
value memories SP1, SP2, SP3 and SP4.
Press the P key to open the respective set value
memory.
The selected set value memory, e.g. SP3 flashes
on the upper PV display.
The corresponding current set temperature
appears on the lower SV display.
Fig. 33: Set value memory SP3 entry
Press the 5 key to increase the set
temperature.
Press the  key to decrease the set
temperature.
Press the 5 and  key to raise and
lower the value by 0.1 respectively.
If the keys are held pressed for at
least one second, the value increases
or decreases quickly and after two
seconds even more quickly; this
means the desired value can be
reached rapidly.
Press the P key to confirm the set temperature.
The set value memory closes and the display
returns to the parameter level.
Press and hold the  or 5 key to return to the
calibration mode.
If no key is pressed for approx. 15
seconds, the device automatically
returns to a previous level up to the
calibration mode.
6.3.6 Retrieving the saved set temperatures
The saved set temperatures can be retrieved in
calibration mode.
Press the U key for approx 2 sec., the current set
value memory opens.
The current reference temperature appears on
the upper PV display.
The set value memory SP1, SP2, SP3 or SP4
appears on the lower SV display for 2 sec.
followed by the current set temperature.
GB
19