6.3.7.1 Setting the heating gradient
GB
The heating gradient "SLor" is active if the refer-
ence temperature is lower than the set temperature.
The setting range extends from 99.99 °C/min up to
0.00 °C/min.
The function is deactivated if SLor
= InF (In no Function) has been set.
You are in the parameter level (as described in
section 5.3.7).
'rEG appears on the upper PV display.
SLor flashes on the lower SV display.
Fig. 40: Parameters for heating gradient SLor
Press the P key.
SLor flashes on the upper PV display.
The respective currently set heating gradient
appears on the lower SV display.
Fig. 41: Heating gradient entry
Press the 5 key to increase the heating gradient
SLor.
Press the  key to decrease the heating
gradient SLor.
Press the 5 and  key to raise and
lower the value by 0.1 respectively.
If the keys are held pressed for at
least one second, the value increases
or decreases quickly and after two
seconds even more quickly; this
means the desired value can be
reached rapidly.
22
WIKA Operating Instructions Temperature Dry Well Calibrators/Micro Calibration Baths
Press the P key to confirm the set heating
gradient SLor.
The display returns to the parameter level and
you can set the other parameters.
If no key is pressed for approx. 15
seconds, the device automatically
returns to a previous level up to the
calibration mode.
After carrying out the settings, the
calibrator uses the new values only
when modifying the set temperature
or switching the calibrator/micro
calibration bath off and on again.
6.3.7.2 Setting the cooling gradient
The cooling gradient "SLoF" is active if the
reference temperature is higher than the set
temperature.
The setting range extends from 99.99 °C/min up to
0.00 °C/min.
The function is deactivated if SLoF
= InF (In no Function) has been set.
You are in the parameter level (as described in
section 5.3.7).
'rEG appears on the upper PV display.
SLor flashes on the lower SV display.
Fig. 42: Parameters for heating gradient SLor
Use the 5 or  key to select the parameter
SLoF.
'rEG appears on the upper PV display.
SLoF flashes on the lower SV display.